StatIM FAQ's
StatIM Cassette Sterilizers
Assorted Questions and Answers provided by SciCan
Q: What type of water do I use with Statim?
A: Always use steam distilled water in the STATIM Cassette
Autoclave. Other water sources will damage the unit and void the warranty.
Q: How do I test the quality of distilled water?
A: If purchasing distilled water, SciCan recommends using a water
conductivity tester (SciCan part no. 01-103139S) to check for impurities. If you
would like to produce your own distilled water, take a look at SciCan's water
distiller product offering.
Q: How do I prepare instruments for use with Statim?
A: Thoroughly wash, rinse, and dry all instruments before inserting into
the Statim.
Q: How often do I clean the interior and exterior of the cassette?
A: Keeping the Statim cassette clean is good clinical practice. The
interior surface of the cassette should be cleaned at least once a week with a
mild dishwashing soap. (For instructions, see the Statim Operators Manual). To
promote quick drying, we recommend spraying the cassette lid and tray with Stat-dri
after each cleaning. After spraying, let the cassette lid and tray dry
completely.
Q: How do I lubricate the rubber cassette seal?
A: Lubricate the rubber cassette seal with a mild liquid soap. In a busy
practice, this should be done 1-2 times per week.
Q: What do I do if steam or a puddle of water is detected during or after
a cycle?
A: If steam or a puddle of water is detected during or after a cycle,
check the cassette seal. This problem may be corrected by doing the following:
- a) reseating the seal
- b) relubricating the seal and probe holes
- c) cleaning the outside edge of the tray
Q: What if I receive an error code?
A: If the LCD displays a NOT STERILE message, note the error code
and consult your Operators Manual.
Q: Do I need to ensure that the Statim is level before operating?
A: Yes. Before operating your Statim autoclave, make sure it rests on a
flat and level surface. If you are uncertain, the level indicator on the front
panel and leveler feet underneath the unit will help you correct any problems.
(Align the bubble in the level indicator to the right front quadrant of the
target.)
Q: What do I do if my Statim has not been run for several days?
A: If you do not operate your Statim autoclave for several days, drain
the reservoir.
Q: How should I drain the reservoir?
A: To empty the reservoir, use the drain tube. If your Statim does not
have a drain tube, a hand pump will suffice. Except when filling or emptying the
reservoir, keep the reservoir cap securely fastened at all times.
Q: What should I do if the Statim is on, but not in use?
A: If your Statim is powered ON but not in use, pull the cassette 1-2cm
out of the unit. This will make the unit inactive until the cassette is
reinserted and will prolong the life of your unit.
Q: How do I recognize whether my filter needs replacement?
A: Monitor the white air filter at the back of the unit. A dirty filter
will cause poor drying. If the filter has turned gray, it is in need of
replacement (see the maintenance section of the Operator's Manual).
Q: How do I transport my Statim?
A: Whenever you need to transport your Statim, make sure you empty the
water reservoir and screw-in the feet. Repack your Statim in the original
packing materials to reduce any damage.
NOTE TO DENTISTS: We are not authorized to ship directly to a dentist office. We generally have better pricing than dental dealers, so we get a lot of dentists hoping to purchase from us, however the only way we can sell (indirectly) to a dentist or oral surgeon is if we have the system shipped to a hospital or surgery center. If you are affiliated with a hospital, we can sell to the hospital and ship there. But we cannot ship directly to a dentist office. No exceptions. Sorry!
NOTE TO CANADIANS: Under no circumstances are we able to ship a SciCan system into Canada. We are only authorized to sell within the United States.
|
Why Fisher?
Family owned and operated for over 25 years, Fisher Biomedical specializes in consulting with clients on their sterilization needs, having worked with medical, dental, and opthalmological offices, surgery centers and hospitals, veterinary practices, colleges and universities, government agencies and military installations, everything from small med spas up through major institutions and manufacturing laboratories — across the USA.
When you call us, you speak with an expert who can answer your questions and help you not only find the right system, at the right price — but also get up and running successfully, rapidly,effectively.
No other seller or service-provider offers this level of expertise and support. Don't settle for second best!
|